While we strive to provide the highest level of service in monitoring conditions, there are certain areas that fall outside the scope of our responsibilities:

  1. Installation of Sensors: Clients are responsible for the proper installation of all sensors used for condition monitoring. Installation should be completed in accordance with the manufacturer’s specifications and Recommended Guidelines.
  2. Handling Freezer Contents: For the safety and integrity of your assets, as well as the well-being of our team, we do not handle, move, or otherwise interact with the items stored within client-maintained environments. Our service is strictly limited to managing technical infrastructure and the configuration and delivery of sensors.
  • Application User Guide This information on the use of the monitoring application

    The web application link to access the system is https://freezermonitor.its.yale.edu. After access being granted, you can login using your Yale NetID and password.

    Please note: The web application is compatible with Chrome, Edge, and Firefox browsers, but it does not work with Safari.

  • Application Custom View and Preferences:
    • For all users - To create a custom view to only see your department sensor or for preferences, please see the instructions HERE.
    • For PI/Responsible Party only - To activate/inactivate a sensor, update alert contacts or change temperature threshold for alerts for a sensor, please see the instructions HERE.
  • Events / Alerts:

    Below is the list of system alerts you will receive if they are triggered. The alerts are configured to send notifications to the alert contacts you designated in three escalations. NOTE:  if the 1st contact(s) dismisses the alert, the secondary and tertiary contacts will not receive any alerts.

    Please note: While we provide timely alerts on sensor thresholds, it is the client’s responsibility to take corrective action based on those alerts. Our service does not include responding to or resolving issues triggered by such notifications. It’s very important to Dismiss (Acknowledge) temperature alert notifications when received. Failure to do so will escalate the alerts to EHS as the final escalation point.

Alert Type
Explanation (When triggered) 
Threshold and Alert Name
Temperature

Temperature out of defined range (for 3 levels of alert contacts)

  1. Temp out of Range - 15 Minutes
  2. Temp out of Range - 30 Minutes
  3. Temp out of Range - 45 Minutes
Asset Properties Dismissing (acknowledging) a temperature alert will stop alert escalation. Once dismissed, the unit will go into Maintenance mode, and the system will trigger a secondary process called Asset Properties. 
  1. Asset Properties – Manual: This will be triggered immediately after acknowledging a temperature alert. This can be acknowledged if the cause of the temperature alert is resolved within 3 hours. And if not acknowledged for any reason…
  2. Asset Properties – Automatic: This will be triggered and acknowledged automatically 3 hours after a temperature alert is acknowledged, resetting it to monitor the temperature again.
Sensor Monitor (Door Alert) Door contacts open/close (for 1st level of alert contacts) Door Open - 5 Minutes.  This alert has 5 minutes reminder
Sensor Tag Error Sensor errors that prevent transmitting data
  1. Sensor Tag Error - 15 Minutes
  2.  Sensor Tag Error - 30 Minutes
  3.  Sensor Tag Error - 45 Minutes
Out of Sight Wi-Fi connectivity issues (for 2 levels of alert contacts)
  1. Out of Sight - 1 Hours
  2. Out of Sight - 2 Hours
External Power External power disconnection - if connected (for 3 levels of alert contacts)
  1. AC Power Loss - 5 Minute
  2. AC Power Loss - 15 Minute
  3. AC Power Loss - 30 Minute
Battery Level Battery level reaches a specified threshold (for 1st level of alert contacts)

Battery Low - 10%

This alert has 24 hours reminder

  • Tag Calibration Program:

    HERE you can find the information that you might need about the Tag Calibration Program. Please contact Securitas directly if you are interested in the program or have any questions about it.